ASHRAE 51-07 PDF

  • June 14, 2019

Standard and ANSI/ASHRAE Standard 51, except as modified by Units shall be tested in accordance with ANSI/AMCA Standard /ANSI/ASHRAE. Standard — Laboratory Methods of Testing Fans for Aerodynamic Performance Rating (ANSI/AMCA Standard ) (ANSI/ASHRAE Approved). and Air-Conditioning Engineers (ASHRAE) approved ANSI/AMCA Standard ANSI/ASHRAE Standard , Laboratory Methods of Testing Fans for.

Author: Ditilar Arashiramar
Country: Saint Kitts and Nevis
Language: English (Spanish)
Genre: Automotive
Published (Last): 28 June 2007
Pages: 207
PDF File Size: 13.38 Mb
ePub File Size: 15.46 Mb
ISBN: 721-5-53841-112-7
Downloads: 82683
Price: Free* [*Free Regsitration Required]
Uploader: Tucage

If a portion of the internal energy is converted into kinetic energy, the static temperature is decreased accordingly. The specification permitting an uncertainty based on the maximum observed test reading during the test leads to combined relative uncertainties in both fan pressure and fan airflow rate that are higher at low values of the fan pressure or fan airflow rate than at high values of those test results.

A torque measurement device shall have a static calibration and may have a running calibration through its range of use. The description of the test fan shall be recorded. During a laboratory test, the air density and rotational speed may vary slightly from one determination point to another.

The minimum distance between the centers of any two nozzles in simultaneous use shall be three times the throat diameter of the larger nozzle. Each pressure tap shall conform to the requirements in Section 4. The procedure is to pressurize or evacuate the test chamber and use a flow meter to establish the leakage flow rate. The coefficient of friction f shall be determined from [19]: A leakage test should be performed prior to initial use and periodically thereafter, with corrective action taken if necessary.

This is generally acceptable because fans are not usually rated at the low pressure or low flow portions of their characteristic curves. Other factors may have an equal or greater effect on fan performance. In each case, the difference between the two readings shall be within 0. The velocity pressure at a point shall be measured on an indicator such as a manometer with one leg open to a total pressure sensor, such as the impact tap of a Pitot-static tube, and the other leg connected to a static pressure sensor such as the static tap of the same Pitot-static tube.

  CANON IR C2550 MANUAL PDF

For the test, positions around the periphery of the jets and slightly upstream of the settling means shall be scanned for reverse velocities.

Groups approve laboratory methods for testing fans.

Tubing longer than 1. This constant is derived by converting to the SI equivalent units: A flow nozzle without an integral throat tap may be used in a multiple nozzle chamber, in which case, upstream and downstream pressure taps shall be located as shown in the figure for the appropriate setup.

The nozzle shall have a cross-section consisting of elliptical and cylindrical portions, as shown. Generally, several screens in each airflow settling means will be required. A total pressure tube is sensitive only to the pressure in the immediate vicinity of the open end. When the fan discharges into a duct with a Pitot traverse, total pressure Pt2 shall be considered equal to the sum of the average static pressure Ps3 and the velocity pressure Pv3 corrected for the friction loss due to both the equivalent length Le of the straightener and the length L2,3 of the duct between the fan outlet and the measurement plane.

A complete set of measurements for a particular point of operation of a fan. The pressure signal can then be transmitted to an indicator. Diameter ratio has been omitted in Section 7. Multi-point or continuous-record averaging can be accomplished with instruments or analyzers designed for this purpose. P Q The fan power output is proportional to the shaded area, which leads to: Efficiency is based on a per-unit basis.

  CALCULO II VICTOR CHUNGARA PDF

WOLTER PTE LTD.

A duct with a measurement plane for airflow determination shall be straight and have a uniform circular cross-section. A pre-test uncertainties analysis is not required for each test covered by this standard because it is recognized that most laboratory tests for rating are conducted in facilities where similar tests are repeatedly run. The uncertainty in the results will be expressed in two parts, both of which will be based on the uncertainties in various measurements. A flow nozzle with an integral throat tap shall be used for a ducted flow nozzle setup.

The contribution to the combined uncertainty in the airflow measurement shall not exceed ashrwe corresponding to ashtae.

Dotted lines on the outlet indicate a diffuser cone which may be used to approach more nearly free delivery. This is easily obtained by using the specified calibration procedures. However, it is possible to perform an uncertainty analysis to identify a range of values within ashhrae the true value probably lies. Static pressure taps shall be in accordance with Figure 2A.

amca 210 07 airflow nozzle

Air temperature measured by a temperature sensing device without modification to compensate for the effect of humidity. Using test values for Q, Pt, and H with Equations 7. Since the air velocity in an inlet chamber is considered uniform due to the settling means employed, a single measurement is representative of the average chamber pressure.

See Annex B for two recommended leakage test methods.